PURE DSA

A medical imaging phantom for the QA of digital subtraction angiography. It contains a moving pneumatic plate with Aluminium strips of varying thicknesses to simulate blood vessels of different sizes. This phantom complies with DIN 6868/parts 8, DIN 6868/4 (2007) and 54.

It allows for evaluation of:

  • Dynamic range
  • Artefacts
  • Contrast media sensitivity
  • Logarithmic errors

Specification:

  • Dynamic Copper step-wedge with steps ranging from 0.20 to 1.40mm
  • Moving pneumatic PMMA plate with a set of 4 Aluminium strips of varying thicknesses (0.05, 0.1, 0.2, 0.4mm) 99.5% purity
  • Pneumatic pump for moving the PMMA plate

Comes with:

  • Detailed user manual
  • Robust carrying case for easy transport

Complies with:

  • DIN 6868/parts 8
  • DIN 6868/4 (2007) and 54